Burning Characteristics Lab Test Machine
Product details:Model ZLT-HV4
The ZLT Horizontal Vertical Flame Chamber Tester is a fully automatic instrument contained in its own cabinet to maximize the safety of the operator. the device complies with UL1581 2001 (1060 FT1,1061,1080VW 1,1090,1100FT2) standards related requirements designed. Applicable to the single wire and cable, and soft line horizontal (FT1), vertical (VW1) combustion test. Burning time and burning the number of adjustable, can be automatic / manual, vertical / horizontal switching, gas pressure, flow adjustable and equipped with flame temperature recognition device.
SPECIFICATIONS
Flame standard |
ANSI/ASTM D 5207 & UL 1581 |
Burner Tube |
ANSI/ASTM D 5025 |
Burning Angle |
0°,20° |
Thermocouple |
Type K (Ni/Cr – Ni/Al),200°~1050° |
Thermocouple Size O/D |
Φ0.5mm,type K |
Temperature Resolution |
1oC |
Standard Copper Block |
weight:10 ± 0 .05 g before drilling. Cu-ETP |
Time for Temperature to Increase
|
1、225W/50mm(16.5mm)/84±2S A:405 ±10mL/min& 45±5 mm water 2、500W/125mm(40mm)/54±2S A:965 ±30mL/min&125±25 mm water B:385 ±15mL/min&552±100 mm water C:300 ±10mL/ min&600±125 mm water |
Gas Supply(do not provide) |
methane,min.98% purity |
Input Power |
220 V 50 Hz 3A or 115 V 60HZ Optional |
Volume Of Chamber |
>4m3, black interior |
Outer Dimensions |
W*D*H=2000mm*1000mm*2000mm |
Conform to standard |
UL1581:1060/1061/1080/1090/1100 |
Includes: Flame height gauge and 100% absorbent cotton